Overview
This document provides essential information on the PL40-I Level Transmitter, including safety precautions, installation guidelines, operating conditions, troubleshooting, and maintenance. The PL40-I is a 2-wire, 4-20mA level transmitter designed for hazardous area applications, ensuring precise level measurements in various tank systems.
1. Safety Information (p.4-5)
- Electrical Hazards: Power isolation is mandatory before installation or maintenance.
- Intrinsic Safety Compliance: Certified for Ex ia IIC T6...T1 Ga (ATEX & IECEx standards).
- Electrostatic Discharge (ESD) Precautions:
- Avoid exposure to static electricity discharges.
- Handle and store electronic components properly.
2. Storage Before Installation (p.6)
- Store in a dry, sheltered area and keep in original packaging.
- Inspect for physical damage before installation.
3. Description and Application (p.9)
- The PL40-I Level Transmitter consists of a transducer and transmitter box connected by a shielded cable.
- Designed for tank level measurements, including ballast, fuel oil, fresh water, and chemical products.
- Suitable for hazardous area installations with intrinsic safety certification.
4. Technical Specifications (p.11-12)
- Measurement Ranges:
- Level measurement: 0 to 70 bar
- Overload capacity: up to 105 bar
- Operating Temperature:
- Transducer: -40°C to +125°C
- Transmitter Box: -40°C to +85°C
- Protection Ratings:
- Transducer: IP68 (fully submersible)
- Transmitter Box: IP66/67 (outdoor-rated)
- Output Signal: 4-20mA DC (2-wire system).
5. Mechanical Installation (p.15-16)
- Mounting Options:
- Immersed mounting for ballast tank applications.
- Side flange mounting with adapter or welded flange.
- Isolation Valve Recommended: Install between tank and transducer for easier maintenance.
6. Electrical Installation (p.16-19)
- Transducer Cable Requirements:
- Must be free from damage (no cracks, burns, or crushing).
- Transmitter Box Placement:
- Install outside the tank in a protected area.
- Avoid exposure to shock, chemicals, and excessive moisture.
- Intrinsic Safety Guidelines:
- Use an isolation barrier (KFD0-CS-Ex1.51P/2.51P) in the safety area.
- Certified for ATEX & IECEx explosion-proof applications.
7. Operation & Calibration (p.19-20)
- Startup Verification:
- 3.5mA output → No signal detected.
- 4-20mA output → Normal level or pressure reading.
- 21mA output → Overpressure alarm.
- Calibration:
- Factory pre-set, no adjustment needed.
- Ensure proper specific gravity setting for accurate readings.
8. Maintenance & Troubleshooting (p.20-21)
- Common Issues & Solutions:
- Indication Error: Check transducer cleanliness, zero/span settings, and venting pipe.
- Zero Reading: Inspect power supply, isolation valve, and sensor status.
- Maximum Reading: Likely sensor failure—requires replacement.
Comments
0 comments
Please sign in to leave a comment.