Overview
This document provides the PL40-U Pressure Transmitter operating procedures, including installation, safety precautions, technical specifications, and maintenance guidelines. The PL40-U is designed for process pressure measurement of liquids, gases, and chemical products, supporting intrinsically safe applications in hazardous areas.
1. Safety Information (p.4-5)
-
Danger & Warning Notices:
- DANGER: Ignoring safety instructions may result in serious injury or death.
- WARNING: Failure to follow guidelines may lead to equipment damage or personal injury.
-
Electrical Safety:
- Ensure the power is disconnected before performing installation or maintenance.
- Obtain the necessary work permits before servicing the device.
-
Anti-Static Precautions:
- Electronic components are electrostatic-sensitive devices (ESD).
- Handle PCBs with proper grounding, and avoid direct contact with circuit components.
2. Storage Before Installation (p.6)
- Packaging: Equipment is packed in heavy-duty cartons, but is not waterproof.
-
Storage Conditions:
- Keep in a dry, covered area, avoiding extreme temperatures.
- Perform pre-installation checks for physical damage, bent connectors, or scratches.
3. Description and Application (p.9)
-
Description:
- The PL40-U Pressure Transmitter measures gauge and absolute pressure in gases, steams, liquids, and dusts.
- Designed for hazardous areas with intrinsically safe certification.
-
Applications:
- Suitable for marine, industrial, and chemical processing environments.
- Supports various process connections including hygienic, threaded, flanges, and diaphragm seals.
4. Technical Specifications (p.11-13)
4.1 Measurement Ranges
-
Gauge/Absolute Pressure Measurement:
- 0~100mbar (4bar overload)
- 0~10bar (40bar overload)
- 0~70bar (105bar overload)
4.2 Operating Conditions
- PL40-U-O: -40°C to +85°C
- PL40-U-I: -40°C to +85°C
- Protection Class: IP67 (Water & Dust Resistant)
4.3 Electrical Specifications
- Power Supply: 12-28VDC
- Output Signal: 4-20mA, 2-wire system
-
Load Calculation Formula:
- Rmax = (Uloop – 12V) / 0.021A
4.4 Accuracy & Stability
- Reference Accuracy: ±0.20% F.S. at 25°C
- Temperature Drift: ±0.010% (0-70°C)
5. Mechanical Installation (p.13-14)
-
Mounting Guidelines:
- Ensure proper sensor alignment for accurate pressure readings.
- Avoid excessive vibration or direct exposure to high-pressure surges.
6. Electrical Installation (p.14-16)
-
Precautions for Intrinsically Safe Installations (Ex ia):
- Certified ATEX & IECEx compliance for Zone 0, 1, 2 hazardous areas.
- Requires an intrinsically safe power supply (barrier).
- Ensure proper grounding to prevent electrostatic charge hazards.
-
Connection Diagram:
- PL40-U-O: Uses IP67 Stainless Steel housing with pre-installed cable glands.
- PL40-U-I: Direct connection via PCB terminals.
7. Startup & Operation (p.17)
7.1 Verification
- Ensure correct polarity for all connections before powering ON.
7.2 Signal Testing
- 3.5mA: No pressure signal detected.
- 4-20mA: Normal pressure range.
- 21mA: Overpressure detected.
7.3 Calibration
- Factory Pre-Set: No manual calibration required.
- Measurement Range: Clearly indicated on the transmitter housing.
8. Maintenance & Troubleshooting (p.17-18)
8.1 Common Issues & Solutions
Issue | Possible Cause | Solution |
---|---|---|
No Output Signal | Power supply failure | Check power source & fuse. |
3.5mA Output | No pressure signal detected | Verify pressure source & connections. |
21mA Output | Overpressure condition | Inspect system pressure settings. |
Display Error | Zero/Span misalignment | Adjust settings using Zero/Span. |
9. Ordering Information (p.19)
-
Product Variants:
- PL40-U-O (Outdoor Housing)
- PL40-U-I (Indoor Housing)
10. Revision History (p.19-20)
- R10 (2025.02.28): Form revision & content updates.