1. Adjusting the Alarm Unit Settings
Follow these steps to modify settings in the OVMS-50 alarm unit for the oxygen monitoring system:
-
Access the Alarm Settings
- Press the "M" key on the alarm unit.
-
Select the Channel
- Choose Channel 1 and press the Enter key until the channel setting menu appears.
-
Enter Password
- Input the password: "0000" and confirm.
-
Modify High Scale Setting
- Use the Arrow button to navigate.
- Set the High Scale value to 25.0.
-
Return to Operating Mode
- Press the "Go Back" button to exit settings.
2. Configuring the Alarm Values
-
Modify Alarm Thresholds
- Adjust the alarm values according to the required specifications.
-
Save and Exit
- Press the "Go Back" button to return to the operating mode.
-
Configure Channel 2 Settings
- Select Channel 2 and press Enter until the channel setting menu appears.
-
Adjust Required Settings
- Modify parameters as needed.
-
Finalize Settings
- Confirm all changes and press "Go Back" multiple times to exit to the main menu.
3. Verifying the Alarm System
- Check the Oxygen High Alarm Functionality
- Confirm the Non-Flow Alarm Settings
- Inspect the Vapor Pressure Alarm Configuration
If the alarm still malfunctions, consider replacing the alarm unit with a new one.
4. Final Notes
- Ensure correct wiring and secure cable connections before adjusting the settings.
- If issues persist after adjusting the oxygen monitoring system, verify the sensor and signal input values.